Model | TH9010 | TH9010A |
Number of Units | Eight separate channel | Four separate channel |
Withstanding Voltage Test | |
Output Voltage | AC | 0.10kV 5.00kV 2% |
DC | 0.10kV 6.00kV 2% |
Current TestRange | AC | 0mA 10.00mA (2% readings + 5 digits) |
DC | 0uA 5.00mA (2% readings + 5 digits) |
Rapid Discharge Function | Discharge after test ends (DCW) |
Insulation Resistance Test | |
Output Voltage | 0.10kV 1.00kV 2% |
Resistance Test Range | 0.1M 10.0G |
Resistance Test Accuracy | 0.10M 999M 10% reading+5digits 1.00G 10.0G 20% reading+5digits |
Discharge Function | Discharge after test ends |
Arc Detection | |
Test Range | Corresponding Current | 1mA 20mA |
General Specification | |
Voltage Rising Time | 0.1s 999.9s |
Test time Setting (AC/DC) | 0.2s 999.9s |
Voltage Fall Time | 0.1s 999.9s |
Waiting Time (IR) | 0.2s 999.9s |
Time Accuracy | (1%+0.1s) |
Memory | Store 100 test files and each file can hold at most 20 testing steps |
Interface | |
Standard | HANDLER, RS232, USB DRV, USB HOST |