| Range |
Voltage Measurement |
Resolution |
Accuracy |
| 0.2V |
0|V|0.212V |
10¼V |
±(0.015% + 225¼V) |
| 2V |
0|V|2.12V |
100¼V |
±(0.02% + 350¼V) |
| 20V |
0|V|21.2V |
1mV |
±(0.015% + 5mV) |
| 200V |
0|V|212V |
10mV |
±(0.015% + 50mV) |
| Current Measurement |
| Range |
Current Measurement |
Resolution |
Accuracy |
| 10nA |
0|I|10.6nA |
1pA |
±(0.10 % + 50pA) |
| 100nA |
0|I|106nA |
10pA |
±(0.06% + 100pA) |
| 1¼A |
0|I|1.06¼A |
100pA |
±(0.025% + 500pA) |
| 10¼A |
0|I|10.6¼A |
1nA |
±(0.025% + 1.5nA) |
| 100¼A |
0|I|106¼A |
10nA |
±(0.02% + 25nA) |
| 1mA |
0|I|1.06mA |
100nA |
±(0.02% + 200nA) |
| 10mA |
0|I|10.6mA |
1¼A |
±(0.02% + 2.5¼A) |
| 100mA |
0|I|106mA |
10¼A |
±(0.02% + 20¼A) |
| 1A |
0|I|1.06A |
100¼A |
±(0.03% + 1.5mA) |
| 1.5A |
0|I|1.53A |
±(0.05% + 3.5mA) |
| 3A |
0|I|3.06A |
1mA |
±(0.4% + 7mA) |
| 10A1 |
0|I|10.6A |
±(0.4% + 25mA) |
| »Note¼ |
| 1 For pulse mode, not for DC mode. |
| Pulse source (pulse width is the time from 10% rising edge to 90% falling edge, base level: pulse low level, peak level: pulse high level) |
| Minimum programmable pulse width |
50¼s |
| Pulse width programming resolution |
1¼s |
| Interface |
RS232C, USB HOST, USB DEVICE, LAN, HANDLER |
| Environment & Temperature |
| Environment |
Suitable for indoor equipment |
| Temperature of operating condition |
0C - 55C |
| Humidity of operating condition |
30% - 80% RH¼Ënon-condensing¼ |
| Elevation of operating condition |
0 - 2000 meters¼Ë6561ft¼ |
| Temperature of storage condition |
30C - +70C |
| Humidity of storage condition |
10% - 90% RH¼Ënon-condensing¼ |
| Elevation of storage condition |
0 - 4600 meters¼Ë15092ft¼ |
| Warm-up time after power on |
60minutes |
| General Indicator |
| Power supply |
90 V to 264V,47 Hz to 63Hz, maximum |
| Power consumption |
¼250VA |
| Rack mount |
215mmx132mmx490mm |
| Dimension |
235mmx154mmx530mm |
| Weight |
Approx. 8.5kg (single)/10kg (dual) |
| Safety |
Class I Safety |
| EMC standards |
IEC61326-1/EN61326-1 |
| AS/NZS |
CISPR 11 |
| Insulation resistance |
Under the reference working conditions, the insulation resistance between the power terminals and the shell is not less than 50M; Under humid and hot transportation conditions, the insulation resistance between the power terminals and the shell is not less than 2M. |
| Dielectric strength |
Under the reference working conditions, the power terminals and the shell can withstand the rated voltage of 1.5kV, frequency of 50Hz AC voltage for 1 minute, without breakdown and ying arc phenomenon. |
| Leakage current |
3.5mA |
| Safety certification |
CE, cCSA¼s, C-Tick |